AUTOGRAPH AGS-V Series - Specs
Tabletop Precision Universal Testing Machines
| Model Name | AGS-1NVD/ 2NVD/5NVD |
AGS-10NVD/ 20NVD/50NVD |
AGS- 100NVD |
AGS- 500NVD |
AGS- 1kNVD |
AGS- 5kNVD |
AGS- 10kNVD |
||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| Load Capacity | 1N/2N/5N | 10N/20N/50N | 100N | 500N | 1kN | 5kN | 10kN | ||||||
| Loading Method | Direct high-precision constant-speed strain control method using precision ball screw drive with no backlash | ||||||||||||
| Test Force | Accuracy | High-Precision Type 1/500 | Within ±0.5 % of the displayed test force (within the range of 1/1 to 1/500 of the rated load cell capacity) JIS B7721 class 0.5, EN 10002-2 grade 0.5, ISO 7500-1 class 0.5, BS 1610 class 0.5, complies with ASTM E4 *1) |
||||||||||
| Wide-Range Accuracy Type 1/1000 | Within ±1 % of the displayed test force (within the range of 1/1 to 1/1000 of the rated load cell capacity) JIS B7721 class 1, EN 10002-2 grade 1, ISO 7500-1 class 1, BS 1610 class 1, complies with ASTM E4*1) |
||||||||||||
| Standard Accuracy Type 1/500 | Within ±1 % of the displayed test force (within the range of 1/1 to 1/500 of the rated load cell capacity) JIS B7721 class 1, EN 10002-2 grade 1, ISO 7500-1 class 1, BS 1610 class 1, complies with ASTM E4*1) |
||||||||||||
| Test Force Calibration | Select from automatic test force calibration, tension, or tension and compression. | ||||||||||||
| Crosshead | Speed Range | 0.0005 to 1,500 mm/min (can be set as desired, not in steps) | |||||||||||
| Maximum Return Speed | 1,650 mm/min | ||||||||||||
| Crosshead Speed Accuracy *2) | ±0.1% | ||||||||||||
| Crosshead Speed and Allowable Test Force | Up to the load cell capacity for all speeds | ||||||||||||
| Distance between Crosshead and Table (Tensile stroke) *3) | - | 1,180 mm (945 mm, SCG) |
1,180 mm (920 mm, SCG) |
1,180 mm (915 mm, SCG) |
1,180 mm (910 mm, SCG) |
1,180 mm (760 mm, SCG) |
1,180 mm (740 mm, SCG) |
||||||
| Effective Test Width | 425mm | ||||||||||||
| Crosshead Position Detection | Measurement Method | Optical encoder | |||||||||||
| Display Method | Digital display (display resolution: 0.001 mm) | ||||||||||||
| Positional Accuracy | Within ±0.1 % or within ±0.01 mm of indicated value, whichever is greater | ||||||||||||
| Sampling Rate | Maximum 0.2 msec | ||||||||||||
| Test Condition Files | 40 files | ||||||||||||
| Standard Functions |
|
||||||||||||
| Accessories | Load Cell | text text text text text text | text text text text text text | text text text text text text | text text text text text text | text text text text text text | text text text text text text | text text text text text text | |||||
| Other | Power supply cable (2.4 m), rotation rod, DVD media (manual and other electronic data), Safety precautions sheet, Operation controller, Safety glasses, Safety gloves | ||||||||||||
| Size, Weight |
Standard Model | W657 mm×D525 mm×H1,562 mm Approx. 80 kg (100 V), 75 kg (200 V) |
|||||||||||
| 250 mm Extension Model | W657 mm×D525 mm×H1,812 mm Approx. 85 kg (100 V), 80 kg (200 V) |
||||||||||||
| 500 mm Extension Model | W657 mm×D525 mm×H2,062 mm Approx. 90 kg (100 V), 85 kg (200 V) |
||||||||||||
| Power Supply Requirements | 1φ 100 to 115 V (100 V), 1φ 200 to 230 V (200 V) 50/60 Hz 900 VA (100 V), 1.2 kVA (200 V) Power supply voltage fluctuations shall be within ±10 % of the installation value, and type D grounding (100 Ω or less) shall be provided. |
||||||||||||
| Power Consumption | 100 V:650 W、200 V:600 W | ||||||||||||