Autograph AGS-X Series - Specs
AUTOGRAPH Table-TOP Precision Universal Tester
AGS-10kNXD
Specifications Table-top
| Max. Load Capacity | 10kN | 
|---|---|
| Lineup of Models Differing byLoad Cell Capacity | 1N /2N /5N /10N /20N /50N | 
| Loading Method | Direct high-precision constant speed strain control method via non-backlash ball-screw drive | 
| Force:Accuracy : High-Precision Type (1/500, ±0.5%) | Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Measurement : Accuracy : Standard-Precision Type (1/500, ±1%) | Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Measurement : Calibration | Automatic test force calibration: select tensile, compression, or | 
| Crosshead : Speed Range | 0.001 to 1000 mm/min (stepless) | 
| Crosshead : Max. Return Speed | 1500mm/min | 
| Crosshead Speed Accuracy*2 | ±0.1 | 
| Crosshead Speed and Permitted Test Force | To load cell capacity across entire speed range | 
| Crosshead – Table Distance(Tensile stroke) *3 | 1200mm | 
| Effective Test Width | 425mm | 
| Crosshead Position Detection:Measurement Method | Optical encoder | 
| Crosshead Position Detection:Display Method | Digital display (display resolution: 0.001 mm) | 
| Crosshead Position Detection:Positional Accuracy | ±0.1% indicated value or ±0.01 mm, whichever is larger | 
| Sampling Speed | 1000Hz max. *4 | 
| Standard Functions | • Automatic reading of load cell characteristic values | 
| Standard Functions | • External analog input (2 channels) *4 | 
| Standard Functions | • Analog recorder (option) output | 
| Accessories | Load cell (with CAL cable), Power cable (2.5 m),turning rod, cable clamps, instruction manual | 
| Dimensions | W653×D520× | 
| Weight | 85kg | 
| Power Requirements | Single phase 100/120/220/240 V AC | 
| Power Requirements | Supply voltage fluctuations within ±10% of the set value. | 
| Operating Environment | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) | 
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
*  Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
*  Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-50kNXD
Specifications Table-top
| Max. Load Capacity | 50kN | 
|---|---|
| Lineup of Models Differing byLoad Cell Capacity | 50kN | 
| Loading Method | Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive | 
| Force : Accuracy : High-Precision Type (1/500, ±0.5%) | Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Measurement : Accuracy : Standard-Precision Type (1/500, ±1%) | Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Measurement : Calibration | Automatic test force calibration: select tensile, compression, or | 
| Crosshead : Speed Range | 0.001 to 800 mm/min (stepless) | 
| Crosshead : Max. Return Speed | 1100mm/min | 
| Crosshead Speed Accuracy*2 | ±0.1 | 
| Crosshead Speed and Permitted Test Force | To load cell capacity across entire speed range | 
| Crosshead – Table Distance(Tensile stroke) *3 | 1210mm | 
| Effective Test Width | 425mm | 
| Crosshead Position Detection : Measurement Method | Optical encoder | 
| Crosshead Position Detection : Display Method | Digital display (display resolution: 0.001 mm) | 
| Crosshead Position Detection : Positional Accuracy | ±0.1% indicated value or ±0.01 mm, whichever is larger | 
| Sampling Speed | 1000Hz max. *4 | 
| Standard Functions | • Automatic reading of load cell characteristic values | 
| Standard Functions | • External analog input (2 channels) *4 | 
| Standard Functions | • Analog recorder (option) output | 
| Accessories | Load Cell (with CAL cable), Power cable (5 m), turning rod, cable clamps, instruction manual | 
| Dimensions | W718×D641× | 
| Weight | 260kg | 
| Power Requirements | Single phase AC200-230V 50/60Hz 4.0kVA | 
| Power Requirements | Supply voltage fluctuations within ±10% of the set value. | 
| Operating Environment | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) | 
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
*  Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
*  Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-100kNX
Specifications Floor Type
| Capacity | 100kN | 
|---|---|
| Loading Method | Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive | 
| Force Measurement : Accuracy : High-Precision Type(1/500, ±0.5%) | Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Force Measurement : Accuracy : Standard-Precision Type(1/500, ±1%) | Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Force Measurement : Calibration | Automatic test force calibration: select tensile, compression, or tensile and compression | 
| Crosshead : Speed Range | 0.001 to 800 mm/min (stepless) | 
| Crosshead : Max. Return Speed | 1100mm/min | 
| Crosshead Speed Accuracy *2 | ±0.1% | 
| Crosshead Speed and Permitted Test Force | To load cell capacity across entire speed range | 
| Crosshead - Table Distance (Tensile stroke) *3 | 1255mm (745mm, MWG) | 
| Effective Test Width | 600mm | 
| Crosshead Position Detection : Measurement Method | Optical encoder | 
| Crosshead Position Detection : Display Method | Digital display (display resolution: 0.001 mm) | 
| Crosshead Position Detection : Positional Accuracy | ±0.1% indicated value or ±0.01 mm, whichever is larger | 
| Sampling Speed | 1000Hz max. *4 | 
| Test Method Files | 40 files (PC link: 20 files, standalone controller: 20 files) | 
| Standard Functions | • Automatic reading of load cell characteristic values | 
| Standard Functions | • External analog input (2 channels) *4 | 
| Standard Functions | • External digital input (2 channels) *4 | 
| Accessories | Load cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual | 
| Dimensions | W945×D725×H2164mm | 
| Weight | 525kg | 
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
*  Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
*  Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-300kNX
Specifications Floor Type
| Capacity | 300kN | 
|---|---|
| Loading Method | Direct, high-precision, constant-rate strain control using non-backlash precision ball-screw drive | 
| Force Measurement : Accuracy : High-Precision Type(1/500, ±0.5%) | Within ±0.5% indicated test force (at 1/250 to 1/1 load cell rating)  | 
| Force Measurement : Accuracy : Standard-Precision Type(1/500, ±1%) | Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Force Measurement : Calibration | Automatic test force calibration: select tensile, compression, or tensile and compression | 
| Crosshead : Speed Range | 0.001 to 500 mm/min (stepless) | 
| Crosshead : Max. Return Speed | 550mm/min | 
| Crosshead Speed Accuracy *2 | ±0.1% | 
| Crosshead Speed and Permitted Test Force | 0.001 to 200kN:0.001 to 500mm/min(entire speed range) 200kN to 300kN:0.001 to 400mm/min | 
| Crosshead - Table Distance (Tensile stroke) *3 | 1475mm (635mm, MWG) | 
| Effective Test Width | 600mm | 
| Crosshead Position Detection : Measurement Method | Optical encoder | 
| Crosshead Position Detection : Display Method | Digital display (display resolution: 0.001 mm) | 
| Crosshead Position Detection : Positional Accuracy | ±0.1% indicated value or ±0.01 mm, whichever is larger | 
| Sampling Speed | 1000Hz max. *4 | 
| Test Method Files | 40 files (PC link: 20 files, standalone controller: 20 files) | 
| Standard Functions | • Automatic reading of load cell characteristic values | 
| Standard Functions | • External analog input (2 channels) *4 | 
| Standard Functions | • External digital input (2 channels) *4 | 
| Accessories | Load cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual | 
| Dimensions | W945×D725×H2414mm | 
| Weight | 675kg | 
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
*  Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
*  Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-100kNX
Specifications Floor Type
| Power Requirements : 400V model | Three phases AC380-440V | 
|---|---|
| Operating Environment : 400V model | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) | 
| Power Requirements : 200V model | A three phase AC200-230V | 
| Operating Environment : 200V model | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) | 
| Power Requirements : 400V model | Three phases | 
| Operating Environment : 400V model | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) | 
| Power Requirements : 200V model | A three phase | 
| Operating Environment : 200V model | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) | 
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
*  Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
*  Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation
AGS-20kNXD
Specifications Table-top
| Max. Load Capacity | 20kN | 
|---|---|
| Lineup of Models Differing byLoad Cell Capacity | 20kN | 
| Loading Method | Direct high-precision constant speed strain control method via non-backlash ball-screw drive | 
| Force:Accuracy : High-Precision Type (1/500, ±0.5%) | Within ±0.5% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Measurement : Accuracy:Standard-Precision Type (1/500, ±1%) | Within ±1% indicated test force (at 1/500 to 1/1 load cell rating) | 
| Measurement : Calibration | Automatic test force calibration: select tensile, compression, or | 
| Crosshead : Speed Range | 0.001 to 1600 mm/min (stepless) | 
| Crosshead : Max. Return Speed | 2200mm/min | 
| Crosshead Speed Accuracy*2 | ±0.1 | 
| Crosshead Speed and Permitted Test Force | To load cell capacity across entire speed range | 
| Crosshead – Table Distance(Tensile stroke) *3 | 1250mm | 
| Effective Test Width | 425mm | 
| Crosshead Position Detection:Measurement Method | Optical encoder | 
| Crosshead Position Detection:Display Method | Digital display (display resolution: 0.001 mm) | 
| Crosshead Position Detection:Positional Accuracy | ±0.1% indicated value or ±0.01 mm, whichever is larger | 
| Sampling Speed | 1000Hz max. *4 | 
| Standard Functions | • Automatic reading of load cell characteristic values | 
| Standard Functions | • External analog input (2 channels) *4 | 
| Standard Functions | • Analog recorder (option) output | 
| Accessories | Load Cell (with CAL cable), Power cable (5 m),turning rod, cable clamps, instruction manual | 
| Dimensions | W718×D641× | 
| Weight | 235kg | 
| Power Requirements | Single phase AC200-230V 50/60Hz 4.0kVA | 
| Power Requirements | Supply voltage fluctuations within ±10% of the set value. | 
| Operating Environment | Temperature: 5 °C to 40 °C; Humidity: 20% to 80% (no condensation) | 
*1 Official certification after installation is recommended to comply with EN 10002-2, ISO 7500-1, and ASTM E4 standards, JIS B7721.
*2 Crosshead speed accuracy is calculated from the crosshead travel within a prescribed time at a constant speed between 0.5 mm/minute and 500 mm/minute.
*3 The tensile stroke is the effective stroke when SCG (screw-type flat grips) or MWG (non-shift wedge-type grips) are mounted.
*4 TRAPEZIUM X or TRAPEZIUM LITE X is needed for these functions. Moreover, when automatic test force/stress control (auto tuning) and the automatic strain control (auto tuning) are used, the sampling speed becomes 10msec.
*5 Dataletty (option) and TRAPEZIUM X / TRAPEZIUM LITE X can not be used together.
*  Values stated in this catalog are based on measurements conducted according to separately defined inspection standards.
*  Windows7/ Windows Vista/ Windows XP is a registered trademark in the United States and other countries of Microsoft Corporation