EZ Test Specs

Specs

 
Name EZ Test
EZ-SX EZ-LX EZ-LX HS
Tester Load Capacity Max. 500 N Max. 5 kN Max. 2 kN
The load cell type can be selected from 9 types: 1 N, 2 N, 5 N, 10 N, 20 N, 50 N, 100 N, 200 N, or 500 N. The load cell type can be selected from 12 types: 1 N, 2 N, 5 N, 10 N, 20 N, 50 N, 100 N, 200 N, 500 N, 1 kN, 2 kN or 5 kN. (Up to 2 kN for EZ-LX HS)
Load Method High-precision constant-speed strain measurement using backlash-free ball screw drive
Test Force Measurement Load High-Precision Type (note 2) ±0.5% of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4.
Standard-Precision Type (note 2) ±1% of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4.
Range 1 range (rangeless)
Test Force Calibration Automatic calibration using calibration cable
Crosshead Speed Range 0.001 to 1000 mm/min 0.001 to 2000 mm/min
Maximum Return Speed 1500 mm/min 3000 mm/min
Crosshead Speed Accuracy (Note3) Within ±0.1% of test speed
Crosshead Speed and Allowable Test Force Up to the capacity of the load cell at all speeds
Distance Between Crosshead and Jig Mounting Surface 500mm 920mm
Maximum Grip Space 395 mm (500 N max. load cell + tensile jig) 700 mm (5 kN load cell + 5 kN screw type flat grips)
755 mm (1 kN load cell + 1 kN screw type flat grips)
860 mm (500 N max. load cell + tensile jig)
Depth of Test Space 100mm (table section)
Crosshead Position Detection Measurement & Display Optical encoder measurement, digital display (display resolution: 1 µm)
Accuracy ±0.1% of indicated value or ±0.01 mm, whichever is greater
Crosshead Control Single test control (single-direction tension or compression test), cycle test control (repetitive tension or compression test)
Sampling Speed 1 ms max. (TRAPEZIUMX-V/TRAPEZIUM LITE X is needed for this function)
Standard Functions Included Constant test force (creep) control (note 4)
Auto-stop and auto-return functions when specimen fracture is detected (crosshead auto home-position return)
Test condition le function, adjustable crosshead speed function
Display function: Actual test force display or stress display (set by user)
Crosshead displacement display in mm or %/GL (set by user)
Peak point test force and stroke
Test force and displacement analog output: 0 V to 5 V DC full scale, respectively (for external recorder)
USB interface
Manual crosshead position fine adjustment
Adjustable controller
Touch load alarm
Dimensions and Weight W400 × D530 × H885 mm, approx. 33 kg W400 × D530 × H1315 mm, approx. 55 kg
Input Power Supply Voltage (Note 5) Single phase, 100 V to 150 V AC, 50/60 Hz, or 200V to 230V AC, 50/60 Hz
Power Capacity 700 VA 850VA
Installation Environmental Conditions Temperature: 5℃ to 40℃, Humidity: 20% to 80% (no condensation)
Power voltage fluctuation: within ±10%, Vibration: 10 Hz max., Amplitude: 5 µm max.

Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends inspection at an installation site that meets the requirements specified in these standards.

Note 3: Crosshead speed accuracy is calculated from the crosshead displacement within a prescribed time at a constant speed between 0.5 mm/min and 500 mm/min

Note 4: The test force is kept constant at 70% or less of the tester load capacity, for up to 12 hours.

Note 5: Ground resistance should be 100 Ω or less.

EZ-SX

Tester Load Capacity (Note 1) Max. 500 N
Crosshead Speed Range 0.001 to 1000 mm/min
Maximum Return Speed 1500 mm/min
Test Force Measurement : High-Precision Type(Note 2) ±0.5 % of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4.
Test Force Measurement : Standard-Precision Type(Note 2) ±1 % of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4.
Test Force Measurement : Range 1 range (rangeless)
Test Force Measurement : Test Force Calibration Automatic calibration using calibration cable
Dimentions W400 x D530 x H885 mm
Weight Approx. 33 kg

Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.

EZ-LX

Tester Load Capacity (Note 1) Max. 5 kN
Crosshead Speed Range 0.001 to 1000 mm/min
Maximum Return Speed 1500 mm/min
Test Force Measurement : High-Precision Type(Note 2) ±0.5 % of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4.
Test Force Measurement : Standard-Precision Type(Note 2) ±1 % of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4.
Test Force Measurement : Range 1 range (rangeless)
Test Force Measurement : Test Force Calibration Automatic calibration using calibration cable
Dimentions W400 x D530 x H1315 mm
Weight Approx. 55 kg

Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.

EZ-LX HS

Tester Load Capacity (Note 1) Max. 2 kN
Crosshead Speed Range 0.001 to 2000 mm/min
Maximum Return Speed 3000 mm/min
Test Force Measurement : High-Precision Type(Note 2) ±0.5 % of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4.
Test Force Measurement : Standard-Precision Type(Note 2) ±1 % of indicated value (within 1/500 to 1/1 of load cell rated capacity)
Conforms to JIS B 7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4.
Test Force Measurement : Range 1 range (rangeless)
Test Force Measurement : Test Force Calibration Automatic calibration using calibration cable
Dimentions W400 x D530 x H1315 mm
Weight Approx. 55 kg

Note 1: When the load cell capacity is smaller than the tester load capacity, the former is the maximum test force.
Note 2: Shimadzu recommends validation at an installation site that meets the requirements specified in these standards.

EZ-SX

 
Load Capacity 500N
Test Speed 0.001 to 1,000 mm/mim
Return Speed 1,500 mm/min
Test force measurement accuracy : High precision type ±0.5 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4
Test force measurement accuracy : Standard precision type ±1 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4

EZ-LX

 
Load Capacity 5 kN
Test Speed 0.001 to 1,000 mm/mim
Return Speed 1,500 mm/min
Test force measurement accuracy : High precision type ±0.5 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4
Test force measurement accuracy : Standard precision type ±1 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4

EZ-LX HS

 
Load Capacity 2 kN
Test Speed 0.001 to 2,000 mm/mim
Return Speed 3,000 mm/min
Test force measurement accuracy : High precision type ±0.5 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 0.5, ISO 7500-1 class 0.5, EN 10002-2 grade 0.5, and ASTM E4
Test force measurement accuracy : Standard precision type ±1 % of indicated value (Range from 1/500 to 1/1 of the load cell capacity) Complies with JIS B7721 class 1, ISO 7500-1 class 1, EN 10002-2 grade 1, and ASTM E4

For Research Use Only. Not for use in diagnostic procedures. Please contact us to check the availability of these products in your country.