- Specs

Add Bookmark

Product Specification

Measurement Items TOC
TOC Measurement Principle UV oxidation-conductivity measurement method
Measurement Range*1 0 to 500 μg/L
Detection Limit*1 0.02 μg/L
Repeatability*1 ±0.05 μg/L (when concentration is no greater than 1 μg/L)
Accuracy*1 ±5% (500 μg/L sucrose)
Linearity*1 R2 = 0.98 or greater (when measuring 250 μg/L in the 0-500 μg/L range)
Startup Time 5 minutes
Measurement Cycle 5 minutes, 10 minutes, 15 minutes, 30 minutes
External Dimensions 270 (W) × 180 (H) × 140 (D) mm
Required Power Supply 100 to 240 V AC, 50/60 Hz 100 VA
Weight 3.0 kg
Protection From Dust And Water Spray IP33
Wiring Length 10 m or less (when using LAN, analog, and contact)
  • *1 The specifications apply when the oxidation time is set to 70 seconds.

 

Sample Conditions And Installation Environment*3

Sample Temperature*3 25±5 °C (Guaranteed operation: 10 to 50 °C)
Sample Conductivity 0.1 μS/cm or less (Guaranteed operation: 1 μS/cm or less) (both at 25 °C sample temperature)
Sample Flow Rate 30 to 200 mL/min (Recommended: 100 mL/min, guaranteed operation: 30 to 500 mL/min)
Ambient Temperature*3/Humidity

25±5 °C (Guaranteed operation: 10 to 40 °C)
0 to 80 % (No condensation)

  • *2 The samples and environmental conditions must be within the specifications and the changes should be small.
    *3 The difference between the ambient temperature and the sample temperature should be small.

 

{"title":"Specification Sheet Downloads","description":"Download the latest specification sheets.","source":"product","key":0,"max":30,"filter_types":["spec_sheet"],"link_title":"View other Downloads","link_url":""} {"title":"Downloads","description":"Download the latest brochure.","source":"product","key":0,"max":30,"filter_types":["brochures"],"link_title":"View other Downloads","link_url":"","pdf_links":[]}