i-Series - Specs
High Performance Liquid Chromatograph

LC-2070 Series
| Model | LC-2070 ( UV model without sample cooler ) |
LC-2070C (UV model) |
LC-2070C 3D (PDA model) |
|
|---|---|---|---|---|
| P/N | 228-65880-58 | 228-65881-58 | 228-65882-58 | |
| Dimensions | W410 × H605 × D500 mm (Not including reservoir tray height) | |||
| Weight | 58 kg | 63 kg | ||
| Available pH range | 1 to 13 | |||
| Pump | Degassing unit | Five Lines: Mobile phase 4 + Rinse solution 1 (Volume 400 mμL) | ||
| Flow rate setting range | 0.0001 to 10 mL/min | |||
| Configuration | Four-solvent low-pressure gradient | |||
| Gradient/range of set concentrations | 0 to 100%, in 0.1% steps | |||
| Maximum pressure | 50 MPa (0.0001 to 3 mL/min) 44 MPa (3.0001 to 5 mL/min) 22 MPa (5.0001 to 10 mL/min) |
|||
| System Delay Volume | 650 μL (Option: 460 μL, 1100 μL) |
|||
| Pump | Injection method | Total-volume sample injection | ||
| Injection volume accuracy | ±1% (50 L, N = 10) | |||
| Injection volume setting range | 0.1 to 100 μL (Option: 0.1 to 50 μL, 1 to 500 μL, 1 to 2,000 μL) |
|||
| Injection volume reproducibility | RSD < 0.20% (5.0−2000 L) RSD < 0.25% (2.0−4.9 L) RSD < 0.5% (1.0−1.9 L) RSD < 1.0% (0.5−0.9 L) |
|||
| Sample cooler | Not included | 4 to 45 °C | ||
| Detector | UV | UV | PDA | |
LC-2080 Series
| Model | LC-2080C (UV model) |
LC-2080C 3D (PDA model) |
|
|---|---|---|---|
| P/N | 228-65883-58 | 228-65884-58 | |
| Dimensions | W410 × H605 × D500 mm (Not including reservoir tray height) | ||
| Weight | 63 kg | ||
| Available pH range | 1 to 13 | ||
| Pump | Degassing unit | Five Lines: Mobile phase 4 + Rinse solution 1 (Volume 400 mμL) | |
| Flow rate setting range | 0.0001 to 10 mL/min | ||
| Configuration | Four-solvent low-pressure gradient | ||
| Gradient/range of set concentrations | 0 to 100%, in 0.1% steps | ||
| Maximum pressure | 70 MPa (0.0001 to 3 mL/min) 44 MPa (3.0001 to 5 mL/min) 22 MPa (5.0001 to 10 mL/min) |
||
| System Delay Volume | 460 μL (Option: 650 μL, 1100 μL) |
||
| Pump | Injection method | Total-volume sample injection | |
| Injection volume accuracy | ±1% (50 L, N = 10) | ||
| Injection volume setting range | 0.1 to 50 μL (Option: 0.1 to 100 μL, 1 to 500 μL, 1 to 2,000 μL) |
||
| Injection volume reproducibility | RSD < 0.20% (5.0−2000 L) RSD < 0.25% (2.0−4.9 L) RSD < 0.5% (1.0−1.9 L) RSD < 1.0% (0.5−0.9 L) |
||
| Sample cooler | 4 to 45 °C | ||
| Detector | UV | PDA | |
LC-2050C (UV model)
Pump
| Degassing unit |
Five Lines: Mobile phase 4 + Rinse solution 1 [Volume 400 μL] |
|---|---|
| Pumping method |
Parallel-type double plunger |
| Pulsation |
≤ 0.1 MPa [1.0 mL/min,10 MPa, Water] |
| Flow rate setting range |
0.0001 to 10 mL/min |
| Flow rate accuracy |
≤ ±1% or ≤ ±2 μL/min, below whichever is greater |
| Flow rate precision |
< 0.06 %RSD or < 0.02 minSD, below whichever is greater |
| Configuration |
Four-solvent low-pressure gradient |
| Gradient / range of set concentrations |
0 to 100%, in 0.1% steps |
| Gradient / concentration accuracy |
±0.5% [0.1 to 2 mL/min, 1 to 20 MPa, Specified condition] |
| Gradient / concentration precision |
±0.1% [1 mL/min,10 MPa, Specified condition] |
| Maximum pressure |
50 MPa [0.0001 to 3 mL/min] |
| System Delay Volume |
650 μL |
Autosampler
| Sample injection method |
Total-volume sample injection |
|---|---|
| Injection volume accuracy |
±1% [50 μL, N = 10] |
| Injection volume setting range |
0.1 to 100 μL |
| Injection volume reproducibility |
RSD < 0.20% [5.0−2000 μL] |
| Cross-contamination |
0.0025% [Caffeine, Specified condition] |
| Injection cycle time |
Min. 14 sec [Specified condition] |
| Samples for processing |
336 [1 mL], 216 [1.5 mL], 112 [4 mL], 4 [MTP/DWP] |
| Sample cooler |
4 to 45°C |
| Injection linearity |
> 0.9999 |
Column Oven
| Heating and cooling method |
Forced air circulation method |
|---|---|
| Containable column size |
Six columns 10 cm long |
| Temperature control range |
Room temperature −12 °C (−15 °C if an optional unit is installed) to 90 °C |
| Temperature control precision |
±0.1°C |
| Temperature accuracy |
±0.8°C (Specified condition) |
| Flow rate switching valve |
Max. 1 pc |
UV Detector
| Wavelength range |
190 to 700 nm |
|---|---|
| Spectral bandwidth |
8 nm |
| Wavelength accuracy |
≤ ±1 nm |
| Wavelength Reproducibility |
≤ ±0.1 nm |
| Noise |
≤ ±2.5 × 10−6 AU, (250 nm, Specified condition) |
| Drift |
≤ 100 × 10−6 AU/h [250 nm, Specified condition] |
| Simultaneous monitoring of 2 wavelengths |
Enable [Any 2 wavelengths of 190 to 370 nm or 371 to 700 nm] |
| Linearity |
Up to 2.5 AU (5%) |
| Sampling rate |
Up to 100 Hz |
| Light source |
Deuterium (D2) lamp |
| Flow Cell |
12 μL (10 mm, TC), 12 MPa |
| Option cell |
High-Speed: 8 μL [10 mm, TC] |